NewSonic SonoDur 3
NewSonic SonoDur 3
产品价格:¥面议(人民币)
  • 规格:完善
  • 发货地:本地至全国
  • 品牌:
  • 最小起订量:1台
  • 诚信商家
    会员级别:钻石会员
    认证类型:企业认证
    企业证件:通过认证

    商铺名称:深圳市泰立仪器仪表有限公司

    联系人:牟小姐(小姐)

    联系手机:

    固定电话:

    企业邮箱:taili668@vip.163.com

    联系地址:深圳市龙岗区龙翔大道9009号珠江广场A2栋13D室

    邮编:518000

    联系我时,请说是在焊材网上看到的,谢谢!

    商品详情

      品牌:NewSonic 型号:SonoDur 3 加工定制:否
      测量范围:10 – 1999 NV 产品类型:超声波硬度计 试验力:100 N
      允许高度:164x86x23 mm 重量:0.37 kg


      NewSonic SonoDur 3
      The New UCI Generation after SonoDur2

      Mobile Hardness testing onmetals with UCI * in productionand maintenance. Starting withgoods income inspection untilthe fi nal product is reached.

      *Ultrasonic Contact Impedance

      Norms: ASTM A 1038 and DIN 50159-1, 2
      Conversions: ASTM E 140-13, EN ISO 18265

      Multiple Solutions for theSonoDur Family and also forMIC10/MIC20 (by probe retrofi t) Motor probes (HV 0,1 – HV 0,8): 0.1 kgf, 0.3 kgf, 0.8 kgf
      Handheld probes (HV 1 – HV 10): 1 kgf, 3kgf, 5 kgf, 10 kgf

      The better Way to Hardness Testing
      Rugged, easy to use, fast and precise. Made in Germany.

      Instrument Features
      ? Carrying strap, edge protection, IP65, MIL-810G (vibration, shock, drop)
      ? Brilliant color display in 5’’, smooth glass shield for easy cleaning
      ? Direct function access just by touch, intuitive menu control, ideal forMIC10/MIC20 users and others
      ? No set-up needed at beginning of shift – ready for immediate use Android Operating System and Data Transfer
      ? USB: transfer of measurement data in txt-format using thefile explorer
      ? WLAN: automatic transfer of single values, completed data stets
      ? Bluetooth: manual transfer of a set of measurement data
      ?“Unlimited”storage capacity for measurement data and settings

      Intelligent Management ofMeasurement Data and Settings
      ? Simple access to data and qualified back-traceableassessment of results
      ? File names can be predefined, automatic closingof data sets
      ?“AllMeas”- summarized list of results in storeddata sets

      Reliable MeasurementTechnique with long-Term Stability
      Efficient working tool, optimized for the
      daily needs of the inspector

      Unique
      Can be calibrated in full according to
      DAKKS after DIN 50159-2 (calibration
      curve 150 – through 900 HV, diamond,
      nominal test force)

      NewSonic SonoDur 3
      Measuring Specifications
      Measuring principle UCI Method, corresponds to DIN 50159, ASTM A1038
      Test indenter Vickers diamond 136°
      Test loads
      Newton scale (1kgf = 9.81 N)
      Motor probes: 1N (0.1 kgf), 3N (0.3kgf) and 8.6 N (0.9 kgf)
      Handheld Probes: 10N (1 kgf), 30N (3kgf), 49N (5kgf), 98N(10kgf), (Other test loads on request)
      Hardness scales and range
      (according to relevant standards), in this casetable A1 respectively T1, T2 (low alloy steel).
      Different measuring ranges are valid for othermaterials. When exceeding the limits theconversion range will be extended. The calcu-lated values are highlighted in red besides theoriginal data in HV.

      Note:
      Conversions are acc. to latest ASTM
      E140-12b?1(2013) und EN ISO 18265:2014.
      Conversions into tensile strength: 98N (10kgf)test load only.
      Vickers
      Brinell
      Rockwell
      Rockwell
      Rockwell
      Rockwell
      Rockwell    5,6
      Rockwell (EN ISO 18265 only)
      Rockwell
      Knoop (ASTM E140 only)
      Shore (ASTM E140 only)
      Tensile strength
      HV
      HB
      HRB
      HRC
      HRE
      HRF
      HRA
      HRD
      HR45N
      HK
      HS
      MPa
      10 – 1999 (9999)
      76 – 618
      41 – 105
      20,3 – 68
      70 – 108,5
      82,6 – 115,1
      60,7 – 8
      40,3 – 76,9
      19,9 – 75,4
      87 – 920
      34,2 – 97,3
      255 – 2180
      Measurement uncertainty* < 4 % (HV5, HV 10). For other test loads and ranges see table below.
      Relative repeatability* < 5 % (HV5, HV 10). For other test loads and ranges see table below.
      * exceeds DIN 50159, dependent on test load and range (see table below). Specifications are valid for 5 measurements usingVickers reference blocks and according to test conditions given in standard DIN 50159.
      Hardness
      scale
      Measurement uncertainty [%] Relative repeatability [%]
      < 250 HV 250 HV - 500HV 500 HV - 800 HV > 800 HV < 250 HV > 250 HV
      HV 0,1 5 6 7 8 8 6
      HV 0,3 5 6 7 8 8 6
      HV 0,8 4 4 5 6 8 6
      HV 1 4 4 5 6 8 6

      Mechanical and Environmental (Instrument and probe)
      Operating time >10 hours in measurement operation (depending on system performance,temperature and instrument settings), up to 8 hours continuous operation, quick exchangeable battery pack (3.7V 3900mAh LiPolymer)
      Operating Temperature Probe: 0°C to ~ +45°C
      Instrument: -10° ~ +50°C // Charging +10°C ~ +40°C
      Storage Temperature -20°C ~ +70°C
      Humidity Max. 90%, non-condensing
      Dimensions Instrument ca. 164x86x23 mm, Motor probe ?38mm, L=190 mm
      Handheld probe ?25 mm, L=176 mm (free length of rod ca. 12,5 mm)
      Handheld probe ?25 mm, L=207 mm (free length rod ca. 43 mm)
      Weight Instrument ca. 320 gr (incl. battery pack)
      Handheld probe ca. 280 gr, Motor probe ca. 370 gr

      Instrument
      Processor and Memory ARM ® Cortex?-A53 Octa Core 1.3 GHz / System 2GB RAM /
      storage me mory 16 GB eMMC / Micro SD card 4 GB (up to 32 GB )
      Operating system Android 5.1 (Android 7.0)
      Keypad 4 function keys, system touch keyboard
      Power Main battery: 3,7V / 3900mAh, LiPo hard pack, quick exchange
      Charging time: <3h to 80% capacity (Instrument off)
      Shelf Hours: Up to 6 months
      AC Power supply/charger: 90V to 264VAC 50/60Hz to 5VDC
      Display 5“ sunlight readable multi touch display (1280x720 pixel),
      LED-backlight (500 Cd/m2), adjustable
      Interfaces Jack for 5VDC operating/ charging |Docking connector (charging)
      USB 2.0 Micro USB (PC) / probe connector Lemo 4 pos.
      Micro SD-card 4 GB (up to 32 GB) | 2x SIM card
      WLAN 802.11 a/b/g/n | WCDMA/HSDAP/HSPUA, FDD-LTE / TDD-LTE) |
      GSM/GPRS/EDGE (b2/b3/b5/b8 | GPS / AGPS / GLONASS
      Bluetooth 4.0 (supports BLE mode) | NFC | Speaker, microphone
      Sensors / Camera / LED Light sensor, G-sensor, proximity sensor, rear camera 8 mega-pixel,
      multicolor status LED
      IP-Proof IP65 according IEC 60529 Edition 2.1:2001-02
      Drop test MIL-STD-810G Methode 516.6, 4 ft.
      Shock Test MIL-STD-810G Methode 516.6 Prozedur I
      Vibration Test MIL-STD-810G Methode 514.6 Prozedur I
      Instrument Language D, EN, IT, FR, SP, PL, CZ, CN - more on request


      NewSonic SonoDur 3
      Scope of delivery

      SonoDur3, Hardness Tester with Data Logger and Data Export, Data Transfer to PC (USB, WLANor Bluetooth resp.), incl. SONO3-NG, Power Source, SONO2-NG/USB USB-Cable, SONO2-HM, ca. 1,5 mProbe Connection Cable, SONO3-TK-1, Transportation Case, SONO-CD, Product-USB-Stick, SONO3Protect, Protection Foils

      Mandatory accessory
      Motor probe or handheld probe respectively
      Other accessories
      Reference blocks (MPA, company certificate),test stands, guiding harnesses, auxiliary SW.

      Example for configurations
    在线询盘/留言